Logo of The Journal of Bone & Joint Surgery (Br)
Quick search:        
          Advanced Search
Guest Access | Sign In
Journal of Bone and Joint Surgery - British Volume, Vol 91-B, Issue SUPP_III, 392.  
Copyright © 2009 by British Editorial Society of Bone and Joint Surgery
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Google Scholar
Right arrow Articles by Sariali, E.
Right arrow Articles by Fisher, J.
PubMed
Right arrow Articles by Sariali, E.
Right arrow Articles by Fisher, J.

British Society For Computer Aided Orthopaedic Surgery


Glasgow, Scotland: 7–9 February 2008

President: Mr M Maheson


THREE-DIMENSIONAL MODELING OF THE HIP KINEMATICS UNDER MICRO-SEPARATION REGIME AFTER TOTAL HIP REPLACEMENT WITH PARTICULAR REFERENCE TO SQUEAKING

E. Sariali; T. Stewart; Z. Jin; and J. Fisher

Institute of Medical and Biological Engineering, Leeds University, Leeds, UK

The use of hard-on-hard hip prostheses has highlighted specific problems like the "stripe-wear" and the squeaking. Many authors have related these phenomena to a micro-separation between the cup and the head. The goal of the study was to model the hip kinematics under micro-separation regime in order to develop a computational simulator for total hip prosthesis including a joint laxity, and to use it to perform a sound analysis.

A three-dimensional model of the Leeds II hip simulator was developed on ADAMS® software. A spring was used to introduce a controlled micro-separation (less than 500 microns) during the swing phase of the walking cycle. The increase of the load during the stance phase induced a relocation of the head in the cup. Values of the medial-lateral separation predicted from the model were compared to experimental data measured using a LVDT of less than 5 microns precision. Theoretical wear path predicted from the model was compared to the literature data. The frequencies of the vibratory phenomena were determined, using the Fourier transformation.

There was an excellent correlation between the theoretical prediction and the experimental measurement of the medial-lateral separation during the walking cycle (0.92). Edge-loading contact occurred during 57% of the cycle according to the model and 47% according to the experimental data. Velocity and acceleration were increased during the relocation phase in a chaotic manner, leading to vibration. The contact force according to the model had also a chaotic variation during the micro-separation phase, suggesting a chattering movement. Fourier transformation showed many frequencies in the audible area.

A three-dimensional computational model of the kinematics of the hip after total replacement was developed and validated with an excellent precision under micro separation. It highlighted possible explanations for the squeaking that may occur during either relocation phase or edge loading.